VERIFICATION OF FORCE TRANSDUCER FOR DIRECT AND INDIRECT MEASUREMENTS

Abstract

Paper deals with verification of high load force sensor using the small weight weights. Test band was built for this purpose. Verification of test band were executed using the etalon reference sensor. Small forces were executed via using the direct method through the applying of weights. High forces were executed using the indirect method through the lever amplification of load derived from small weights. Uncertainties of measurement were evaluated.

Recommended articles

EFFECTS OF DIFFERENT EXCITATION MECHANISMS IN MACHINE TOOLS WHEN PERFORMING OUTPUT-ONLY MODAL ANALYSIS

MARTIN DIX, MARTIN KOLOUCH, JAN BERTHOLD, JOACHIM REGEL
Keywords: Output-only Modal Analysis | Machine tool | dynamic behaviour | excitation mechanism | white noise